PRODUCTS

Fiber Optic Spectrometers General Purpose Spectrometers

DETAIL

The HS2048 is a compact miniature spectrometer with a configurable spectral range.
It features high resolution (up to 0.07 nm), low stray light (~0.1%), and excellent thermal stability (wavelength drift ~0.1 pixel/°C). It is suitable for laser measurement, plasma emission spectroscopy, color analysis, absorbance measurement, and Raman spectroscopy, making it a high-performance, cost-effective industrial-grade fiber optic spectrometer.



 

  • Features

  • Low stray light: ~0.1%
  • High resolution: up to 0.07 nm
  • Low-temperature drift design, wavelength temperature drift ~0.1pixel/
  • Compatible with Oceanhood multi-core bundle fibers (insertion consistency ≤7%)
  • Strong UV spectral response
  • High performance AD quantization sampling circuit, CCD quantization background noise 30RMS (at minimum integration time);
  • Supports multiple communication interfaces: USB, RS232, and 24PIN interactive interface.
  • Configure proprietary DAC and ADC, can achieve supporting light source enable, intensity control and power feedback.
 

  • Specification

Optical parameters
Optical fiber interface Key-SMA905
Number of pixels 2048pixels
Stray light ~0.1%
Wavelength shift with temperature 0.1pixel/
Consistency of optical fiber insertion and removal 7%
Function parameters
AD sampling 16bit
Data interface USB2.0、RS232
Extended function interface 24PIN
Acquisition mode Single, continuous, software trigger, synchronous external trigger, asynchronous reset external trigger
Detector integration time 60μs-65S
Accuracy of external trigger delay 10ns
CCD read noise1 30
CCD dynamic range2 3000:1
Signal-to-noise ratio 380:1
Response linearity3 98%
Others
Weight 260g
Dimension 98×80×24 mm
Operating temperature 0℃~40℃
Working humidity 20%-85%
1.CCD readout noise: the RMS value of the CCD readout noise at the minimum integration time.
2.CCD dynamic range: calculated under minimum integration time as (saturation value − dark noise baseline) / standard deviation of CCD readout noise, evaluated according to Oceanhood Opto-Electronics standards.
3.Response linearity: The response nonlinearity is measured before calibration.
 

 
  • Products
Model Spectral Range Spectral resolution
Start End
10µm
25µm
50µm 100µm
200µm
HS2048-200-800 200 800 0.90 1.00 1.40 2.40 3.50
HS2048-350-940 350 940 0.80 1.00 1.50 2.40 3.60
HS2048-200-1100 200 1100 1.00 1.80 - 3.00 4.20
HS2048-350-1050 350 1050 0.90 1.00 1.60 2.80 4.00

► Spectral range and parameters can be customized.
► Resolution values may vary by 120% from nominal.


 

  • Dimensional drawing


 

 
  • Testing results

Dark noise (Integral time:10ms)



Dark noise VS integration time



Signal-to-noise ratio



Dynamic range



Standard light source (200-800 nm)



Deuterium tungsten light source (200-800 nm)



Photoluminescence (PL) measurement results