DETAIL

The detector of the spectrometer adopts the CMOS linear image sensor - Hamamatsu S11639 which can provide the spectrometer with high signal-to-noise ratio and large dynamic range. Thanks to the durable integrated circuit and compact design, our spectrometer can provide a better transmission speed and a stable wavelength. It can also achieve stable Raman spectrum measurement from 0℃ to 40℃.

With compact size and superior capability, XS11639 series is able to be applied to optical spectroscopy research, such as absorbance, reflectance, Fluorescence spectrum and Raman spectrum. It can also be used for industrial equipment, such as smoke monitoring, water quality inspection, elemental analysis.

 

  • Features

  • High resolution: resolution<0.6nm
  • Compact and lightweight, suitable for OEM requirement
  • High SNR and large dynamic range
  • Good sensitivity and consistency

  • Specifications

    Model XS11639-630-940
    Detector Hamamatsu S11639
    Spectral range 630-940nm
    Pixels 2048 pixels
    Slit width 25μm (adjustable: 5, 10, 50, 100, 200μm)
    Resolution <0.6nm
    SNR 300:1
    Wavelength accuracy <0.2nm
    Thermal stability <0.01nm/
    A/D resolution 16bit
    Dark noise 30RMS@100ms
    Dynamic range 2000:1@100ms
    Integration time 1ms-65s
    Interface Micro USB
    Dimensions 75×63.5×35mm
    Weight 190g
    Operating temperature 0~40
    Operating humidity 20~85%

 


  • Dimensional drawing




 

  • Applications

  • 633nm & 671nm Raman spectrum
  • Fluorescence measurement
 
 
  • Performance testing: Wavelength accuracy and line width of spectrometer

Test condition:

  • Light source: Ne-1Neon discharge lamp
  • Light fiber105μm
Test result:


  • Performance testing: Response curve of Neon lamp

Test condition:

  • Light sourceNe-1 (Neon discharge lamp)
  • Optical fiber105μm; Integration time: 1ms
Test result:


  • Performance testing: Response curve of Tungsten lamp

Test condition:

  • Light source: Deuterium tungsten lamp
  • Optical fiber: 200μm, integration time: 25ms

Test result:



 

  • Performance testing: Stability of spectral response

Test condition:

  • At a wavelength of 650.784nm of deuterium-tungsten light source, set the integration time for 25ms and preheat for 10 mins
  • The collection was conducted at 2-minute intervals and repeat 100 times
  • Optical fiber: 200μm
Test result:
  • Average value of test: 51278, fluctuating p-p value1924
  • RMS: 0.83%