PRODUCTS

Fiber Optic Spectrometers VIS-NIR Spectrometers

DETAIL
The detector of the spectrometer adopts the CMOS linear image sensor - Hamamatsu S11639 which can provide the spectrometer with high signal-to-noise ratio and large dynamic range.

The spectrometer is able to provide a better transmission speed and a stable wavelength thanks to its durable integrated circuit and compact design. It can also achieve stable Raman spectrum measurement from 0℃ to 40℃.

With compact size and superior capability, XS11639 series is commonly applied to optical spectroscopy research, such as absorbance, reflectance, Fluorescence spectrum and Raman spectrum. It can also be used for industrial equipment, such as smoke monitoring, water quality inspection, elemental analysis.

 

 
  • Features

  • High resolution: resolution<0.4nm
  • Compact and lightweight, suitable for OEM requirement
  • High SNR and large dynamic range
  • Good sensitivity and consistency
 
  • Specifications

    Model XS11639-520-700
    Detector Hamamatsu S11639
    Spectral range 520-700nm
    Pixels 2048 pixels
    Slit width 25μm (adjustable: 5, 10, 50, 100, 200μm)
    Resolution <0.4nm
    SNR 300:1
    Wavelength accuracy <0.2nm
    Thermal stability <0.01nm/
    A/D resolution 16-bit
    Dark noise 30RMS@100ms
    Dynamic range 2000:1@100ms
    Integration time 1ms-65s
    Interface Micro USB
    Dimensions 75×63.5×35mm
    Weight 190g
    Operating temperature 0~40
    Operating humidity 20~85%

 

 
  • Dimensional drawing



  • Applications

  • UV-VIS-NIR absorption spectrum
  • Water quality analysis
  • Fluorescence spectrum
  • HPLC
 
  • Performance testing: Wavelength accuracy and line width of spectrometer

Test condition:

  • Light sourceHG-1-OH (Mercury argon discharge lamp)
  • Light fiber200μm
Test result:


  • Performance testing: Response curve of Neon lamp

Test condition:

  • Light sourceNe-1 (Neon lamp)
  • Optical fiber: 200μm; Integration time: 2ms
Test result:


 

  • Performance testing: Response curve of Deuterium-tungsten lamp

Test condition:

  • Light sourceDeuterium-tungsten lamp
  • Optical fiber: 105μm; Integration time: 1ms
Test result:

  • Performance testing: Spectral response stability

Test condition:

  • At a wavelength of 650.784nm of deuterium-tungsten light source, set the integration time for 25ms and preheat for 10 mins
  • The collection was conducted at 2-minute intervals and repeat 100 times.
  • Optical fiber: 200μm

Test result:

  • Average value: 51278, fluctuating p-p value: 1924
  • RMS: 0.83%