The Area2000 is a compact, back-thinned high-sensitivity miniature spectrometer. It offers configurable spectral range, high resolution (up to 0.07 nm), low stray light (~0.2%), and minimal temperature drift (~0.1 pixel/°C). Ideal for laser measurement, plasma emission spectroscopy, color analysis, absorbance measurement, and Raman spectroscopy, it delivers excellent performance and cost efficiency for industrial applications.
-
Features
- Stray light: ~0.2%
- High resolution: up to 0.07 nm
- Wavelength temperature drift: ~0.1 pixel/°C
- Compatible with Oceanhood multi-core bundle fibers (insertion consistency≤7%)
- Good sensitivity in the UV-VIS range
- CCD background noise≤20 RMS (min. integration time)
- Multiple interfaces including USB, RS232, 24-pin interactive interface
- Built-in DAC and ADC for light source control, intensity adjustment, and power feedback
-
Specification
| Optical parameters | |
| Optical fiber interface | Key-SMA905 |
| Number of pixels | 2048×64 pixels |
| Stray light | ~2‰ |
| Wavelength shift with temperature | 0.1pixel/℃ |
| Consistency of optical fiber insertion and removal | ≤7% |
| Function parameters | |
| AD sampling | 16bit |
| Data interface | USB2.0、RS232 |
| Extended function interface | 24PIN |
| Acquisition mode | Single, continuous, software trigger, synchronous external trigger, asynchronous reset external trigger |
| Detector integration time | 50μs-65S |
| Accuracy of external trigger delay | 10ns |
| CCD read noise1 | ≤20RMS |
| CCD dynamic range2 | 4000:1 |
| Signal-to-noise ratio | 400:1 |
| Response linearity3 | ≥98% |
| Others | |
| Weight | ≈330g |
| Dimension | 100×85×28.5 mm |
| Operating temperature | 0℃~40℃ |
| Working humidity | 20%-85% |
| 1. CCD readout noise: the root mean square (RMS) value of the CCD readout noise at the minimum integration time.; 2. CCD dynamic range: the dynamic range under the minimum integration time, defined as the saturation value divided by the RMS of the CCD readout noise, evaluated according to Oceanhood Opto-Electronics standards; 3. Response linearity: the response nonlinearity is measured before calibration. |
|
-
Products
| Model | Spectral Range | Spectral resolution | |||
| Start | End |
10um
|
25um
|
50um
|
|
|
Area2000-750-1300
|
750 | 1300 | 0.40 | 0.60 | 1.30 |
|
Area2000-200-1100
|
200 | 1100 | 0.80 | 1.30 | 2.00 |
► Spectral range and parameters can be customized.
► Resolution values may vary by 120% from nominal.
-
Dimensional drawing

-
Testing results
Dark noise (Integral time:10ms)Dark noise VS integration time
Signal-to-noise ratio
Dynamic range
Standard light source (200-800 nm)
Deuterium tungsten light source (200-800 nm)
![]()