The detector of the spectrometer adopts the CMOS linear image sensor - Hamamatsu S11639 which can provide the spectrometer with high signal-to-noise ratio and large dynamic range. Thanks to the durable integrated circuit and compact design, XS11639 series would be able to provide a better transmission speed and a stable wavelength. It can also achieve stable Raman spectrum measurement from 0℃ to 40℃.

With compact size and superior capability, XS11639 series can be applied to optical spectroscopy research, such as absorbance, reflectance, Fluorescence spectrum and Raman spectrum.


 

  • Features

  • High resolution: resolution<0.8nm
  • Compact and lightweight, suitable for OEM requirement
  • High SNR and large dynamic range
  • Good sensitivity and consistency

 
  • Specifications

    Model XS11639-790-1050
    Detector Hamamatsu S11639
    Spectral range 790-1050nm
    Pixels 2048 pixels
    Slit width 25μm (adjustable: 5, 10, 50, 100, 200μm)
    Resolution <0.8nm
    SNR 300:1
    Wavelength accuracy <0.2nm
    Thermal stability <0.01nm/
    A/D resolution 16bit
    Dark noise 30RMS@100ms
    Dynamic range 2000:1@100ms
    Integration time 1ms-65s
    Interface Micro USB
    Dimensions 75×63.5×35mm
    Weight 190g
    Operating temperature 0~40
    Operating humidity 20~85%

 
  • Dimensional drawing


 

  • Applications

  • Near infrared absorption
  • Water quality analysis
  • Raman measurement

 
  • Performance testing: Wavelength accuracy and line width of spectrometer

Test condition:

  • Silt wide: 25μm
  • Light sourceHG-1-OH (Mercury argon discharge lamp)
  • Optical fiber200μm
Test result:
 
  • Performance testing: Response curve of Mercury-argon lamp

    Test condition: 
  • Light sourceHG-1-OH (Mercury argon discharge lamp)
  • Optical fiber200μmIntegration time2ms
Test result:
 
  • Performance testing: Response curve of deuterium-tungsten lamp

Test condition:
  • ​​​​​​Light sourceDeuterium tungsten lamp
  • ​​​​Optical fiber105μm; Integration time1ms
Test result:
 
  • Performance testing:  Stability of spectral response

Test condition:
  • At a wavelength of 850nm of deuterium-tungsten light source, set the integration time for 1ms and preheat for 10 mins
  • The collection was conducted at 2-minute intervals and repeat 100 times.
  • Optical fiber: 105μm
Test result:
  • Average value: 45246.55, fluctuating p-p value1398
  • RMS0.64%