CS100 is a miniature fiber optic spectrometer based on symmetrical CT optical path design. It has a compact size and the spectral range can be configured as requested. It features high resolution (the best up to 0.07nm) and low temperature drift(~0.1pixel/), and is suitable for laser measurement, plasma emission spectrum measurement, color measurement, absorbance measurement and Raman measurement and other industrial field measurement. CS100 is an industrial fiber optic spectrometer with high performance.


 

  • Features

  • Based on the traditional CT optical path design, excellent stray light absorption structure and high-performance imported plane reflection grating, the light path stray light ~0.2%;
  • High resolution, the best up to 0.07nm;
  • Low-temperature drift design, wavelength temperature drift ~0.1pixel/;
  • Key-SMA905 design, suitable for kelp pin multi-core dense cluster fiber, fiber insertion strength consistency 7%;
  • Equipped with UV-enhanced linear array CMOS detector, UV spectral response is strong;
  • High performance AD quantization sampling circuit, CCD quantization background noise 30RMS (100ms integration time);
  • Configure USB, serial communication interfaces, configure 24PIN interactive interface, configure proprietary DAC and ADC, can achieve supporting light source enable, intensity control and power feedback.

 
  • Specification

Optical parameters
Optical fiber interface Key-SMA905
Slit 10μm、20μm、25μm、50μm、100μm、200μm
(customizable)
Detector CMOS detector
Number of pixels 2048pixels
Stray light ~0.2%
Wavelength shift with temperature 0.1pixel/
Consistency of optical fiber insertion and removal 7%
Function parameters
AD sampling 16bit
Data interface USB2.0、RS232
Extended function interface 24PIN
Acquisition mode Single, continuous, software trigger, synchronous external trigger, asynchronous reset external trigger
Detector integration time 50μs-65S
CCD read noise1 30
CCD dynamic range2 1300:1
Signal-to-noise ratio 300:1
Response linearity3 98%
Others
Weight 260g
Dimension 98×80×22.5 mm
Operating temperature 0℃~40℃
Working humidity 20%-85%

1. CCD readout noise: integration time 10ms, CCD readout noise root mean square;

2. CCD dynamic range: the dynamic range is the saturation value of 2ms integration time/the root mean square of CCD readout noise, the evaluation method refers to the standard of enterprise;

3. Response linearity: The response nonlinearity is before calibration.
 

 
  • Products
Model Number of pixels Spectral Range Frame rate Spectral resolution
Pixels
Start End frame/s
10um
25um
50um
CS100-200-800
2048
200
800
-
-
0.80
-
CS100-350-950
2048
350
950
-
-
0.80
-


 

  • Dimensional Drawing


  • Testing Results

Dark noise (Integral time:10ms)

 
Dark noise VS Integral time


Signal-to-noise ratio


 
Dynamic range



Standard light source (200-800nm)


 