VIS-NIR Spectrometers

Fiber Optics Spectrometer 660-960nm - XS11639 Series

XS11639-660-960 is a fiber optic spectrometer with a spectral range of 660-960nm.
DETAIL

The detector of the spectrometer adopts the CMOS linear image sensor - Hamamatsu S11639 which can provide the spectrometer with high signal-to-noise ratio and large dynamic range. Besides, XS11639 series can also provide a better transmission speed and a stable wavelength thanks to the durable integrated circuit and compact design. With compact size and superior capability, XS11639 is able to be applied to optical spectroscopy research, such as absorbance, reflectance, Fluorescence spectrum and Raman spectrum.

 

  • Features

  • High resolution: resolution<0.6nm@ 25μm slit
  • Compact and lightweight, suitable for OEM requirement
  • High SNR and large dynamic range
  • Good sensitivity and consistency
 
  • Specifications

    Model XS11639-660-960
    Detector Hamamatsu S11639
    Spectral range 660-960nm
    Pixels 2048 pixels
    Slit width 25μm (adjustable: 5, 10, 50, 100, 200μm)
    Resolution <0.6nm
    SNR 300:1
    Wavelength accuracy <0.2nm
    Thermal stability <0.01nm/
    A/D resolution 16bit
    Dark noise 30RMS@100ms
    Dynamic range 2000:1@100ms
    Integration time 1ms-65s
    Interface Micro USB
    Dimensions 75×63.5×35mm
    Weight 190g
    Operating temperature 0~40
    Operating humidity 20~85%
 
 
  • Dimensional drawing


 

  • Applications

  • Raman spectroscopy
  • Fluorescence measurement
  • Raman measurement
 
 
  • Performance testing: Wavelength accuracy and line width of spectrometer

Test condition:

  • Light sourceNe-1 (Neon discharge lamp)
  • Light fiber105μm
Test result:


  • Performance testing: Response curve of Neon lamp

Test condition

  • Light sourceNe-1 (Neon discharge lamp)
  • Optical fiber105μm; Integration time: 1ms
Test result:

  • Performance testing: Response curve of deuterium-tungsten lamp

Test condition:

  • Light source: Deuterium tungsten lamp
  • Optical fiber: 200μm, integration time: 25ms
Test result:
 
  • Performance testing: Stability of spectral response

Test condition:

  • At a wavelength of 650.784nm of deuterium-tungsten light source, set the integration time for 25ms and preheat for 10 mins
  • The collection was conducted at 2-minute intervals and repeat 100 times.
  • Optical fiber: 200μm
Test result
  • Average value of test: 51278, fluctuating p-p value1924
  • RMS0.83%