-
Features
- Fluorescence spectrum range: 415-1000nm
- Filter cutoff depth: OD5
- High sensitivity: the detection sensitivity is 100 times higher than traditional 90 degrees fluorescence measurement method
-
Dimensional Outline
-
Specifications
Model | Laser wavelength (nm) | Spectrum range (nm) |
FPB-405-1.5-FS | 405 | 415-1000 |
Dimensions |
Main part: 102mm×30mm×13mm
Length of probe: 45mm Length of optical fiber: 115cm±2cm (combined part) / 30cm±2cm (separated part)
|
Laser connector | FC/PC |
Spectrometer connector | SMA905 |
Probe working distance | 7.5mm±1mm |
Fiber core |
Excitation fiber is 105um Collection fiber is 200um
|
The armor size of the combined parts | Φ8.5mm Metal sleeve,Black PVC outer |
Max tension (8.5mm guard tube)
|
50N |
Operating temperature | Probe:-20℃~80℃ |
Humidity | 5%-80% |
Type | Probe with trigger function | Right angle probe | Probe (no tail fiber design) |
Model | FPB-405-1.5T-XX | FPB-405-1.5L -XX | FPB-405-N-XX |
Photo | |||
Laser connector | FC/PC (customizable) | ||
Spectrometer connector | SMA905 (customizable) | ||
Probe working distance | 7.5mm (customizable) |
Raman probe (RPB) Fluorescence probe (FPB) |
+ |
Laser wavelength |
+ |
(+L means right angle probe) (+T means setting the trigger mode) |
+ |
(FC/PC, SMA905…) |
For example: RPB-266-1.5L-FS
RPB: Raman probe
266: Laser wavelength
1.5: Fiber length (1.5m)
L: Right-angle surface probe
F: FC (laser connector)
S: SMA905 (spectrometer connector)
-
Applications
- Drug testing
- Biomedical
- Chemical monitoring
- Gemstone testing and identification
-
Fluorescence Spectrum
405 Fluorescence probe- Ethyl Alcohol
Testing equipment:
Laser:Laser405-1HS
Spectrometer:XS11639-200-850
Probe:FPB-405-1.5-FS-1B
Test condition:
Integration time: 5s
Laser power: 100mw
Test sample: Ethyl Alcohol
405 Fluorescence probe- Acetonitrile
Testing equipment:
Laser:Laser405-1HS
Spectrometer:XS11639-200-850
Probe:FPB-405-1.5-FS-1B
Test condition:
Integration time: 2s
Laser power: 100mw
Test sample: Acetonitrile
405 Fluorescence probe- Cyclohexane
Testing equipment:
Laser:Laser405-1HS
Spectrometer:XS11639-200-850
Probe:FPB-405-1.5-FS-1B
Test condition:
Integration time: 2s
Laser power: 100mw
Test sample: Cyclohexane