XS11071 is a long-focal length, back-illuminated, high-sensitivity fiber optic spectrometer based on symmetrical CT optical path design. It is compact and can be configured within the spectrum range of 200-1100nm. It features high resolution (the best up to 0.1nm) and low stray light (~5‰). Meanwhile, it equipped with high sensitivity back-thinned CCD which has good signal-to-noise ratio and wide dynamic range. It is definitely a valuable fiber optic spectrometer which suitable for laser measurement, plasma emission spectrum measurement, color measurement, absorbance measurement, Raman measurement and other industrial field measurement.
-
Features
- Based on the traditional CT optical path design;
- High resolution, the best up to 0.1nm;
- Key-SMA905 design, suitable for kelp pin multi-core dense cluster fiber, fiber insertion strength consistency ≤7%;
- Equipped with UV enhanced backlit area array CCD detector, UV spectral response is strong, high signal-to-noise ratio;
- High performance AD quantization sampling circuit, CCD quantization background noise ≤10RMS (10ms integration time);
- Configure USB, serial communication interfaces, configure 24PIN interactive interface, configure proprietary DAC and ADC, can achieve supporting light source enable, intensity control and power feedback.
-
Specification
Optical parameters | |
Optical fiber interface | Key-SMA905 |
Slit | 10μm、20μm、25μm、50μm、100μm、200μm (customizable) |
Detector | Back-thinned CCD |
Number of pixels | 2048×64 pixels |
Stray light | ~0.5% |
Consistency of optical fiber insertion and removal | ≤7% |
Function parameters | |
AD sampling | 16bit |
Data interface | USB2.0、RS232 |
Extended function interface | 24PIN |
Acquisition mode | Single, continuous, software trigger, synchronous external trigger, asynchronous reset external trigger |
Detector integration time | 8ms-65S |
CCD read noise1 | ≤10 |
CCD dynamic range2 | 5000:1 |
Signal-to-noise ratio | 550:1 |
Response linearity3 | ≥98% |
Others | |
Part No. | Area |
Weight | ≈1kg |
Dimension | 164×109×56.5 mm |
Operating temperature | 0℃~40℃ |
Working humidity | 20%-85% |
1. CCD readout noise: integration time 10ms, CCD readout noise root mean square; 2. CCD dynamic range: the dynamic range is the saturation value of 2ms integration time/the root mean square of CCD readout noise, the evaluation method refers to the standard of enterprise; 3. Response linearity: The response nonlinearity is before calibration.
|
-
Products
Model | Number of pixels | Spectral Range | Frame rate | Spectral resolution | ||
Pixels
|
Start | End | frame/s |
10um
|
25um
|
|
XS11071-200-1050 |
2048
|
200
|
1050 | - |
-
|
1.00 |
XS11071-535-630 |
2048
|
350
|
950
|
- |
-
|
0.15
|
XS11071-795-1050 | 2048 | 795 | 1050 | - | - | 0.50 |
XS11071-360-800 | 2048 | 360 | 800 | - | - | 0.80 |
-
Dimensional drawing
-
Testing results
Signal-to-noise ratio
Standard light source (200-900nm)