PRODUCTS

Fiber Optic Spectrometers UV-VIS Spectrometers

DETAIL

MS11639 is an industrial miniature optical fiber spectrometer with a spectral range of 200nm~1100nm (configurable). It features high resolution, low stary light and low temperature drift. Further, the detector is using highly sensitive linear array CMOS sensor that can efficiency provide the spectrometer with high signal to noise ratio and wide dynamic range.
 

About the applications, MS11639 can be used in laser measurement, plasma emission spectrum measurement, color measurement, absorbance measurement, Raman measurement and other industrial field measurement.



  • Features

  • Based on the traditional CT optical path design;
  • High resolution, the best up to 0.07nm;
  • Key-SMA905 design, suitable for kelp pin multi-core dense cluster fiber, fiber insertion strength consistency 7%;
  • Patented and unique low-temperature drift design, wavelength temperature drift~0.1 pixel/;
  • Configure USB, serial communication interfaces, configure 24PIN interactive interface, configure proprietary DAC and ADC, can achieve supporting light source enable, intensity control and power feedback.

 
  • Specification

Optical parameters
Optical fiber interface Key-SMA905
Slit 10μm、20μm、25μm、50μm、100μm、200μm
(customizable)
Detector CMOS Detector
Number of pixels 2048 pixels
Stray light ~0.2%
Wavelength shift with temperature ~0.1pixel/
Consistency of optical fiber insertion and removal 7%
Function parameters
AD sampling 16bit
Data interface USB2.0、RS232
Extended function interface 24PIN
Acquisition mode Single, continuous, software trigger, synchronous external trigger, asynchronous reset external trigger
Detector integration time 50μs-65S
CCD read noise1 ≤30
CCD dynamic range2 3500:1
Signal-to-noise ratio 380:1
Response linearity3 98%
Others
Part No. MS2000
Weight ≈630g
Dimension 125×95×35 mm
Operating temperature 0℃~40℃
Working humidity 20%-85%

1. CCD readout noise: integration time 10ms, root mean square of CCD readout noise;

2. CCD dynamic range: the CCD dynamic range is 2ms when the integration time is the saturation value/the root mean square of the CCD readout noise;

3. Response linearity: The response nonlinearity is before calibration.
 

 
  • Products
Model Number of pixels Spectral Range Frame rate Spectral resolution
Pixels
Start End frame/s
10um
25um
MS11639-200-1100
2048
200
1100 - 0.90 1.30


 

  • Dimensional drawing



 
  • Testing results

Dark noise (Integration time:10ms)

 
Dark noise VS Integration time


Signal-to-noise ratio


 
Dynamic range



Standard light source (200-900nm)


 
Deuterium tungsten light source (200-900nm)